How well are heart rates measured by pulse oximeters and electronic sphygmomanometers? Practice-based evidence from an observational study of acutely ill medical patients during hospital admission

How well are heart rates measured by pulse oximeters and electronic sphygmomanometers? Practice-based evidence from an observational study of acutely ill medical patients during hospital admission

Background: heart rates generated by pulse oximeters and electronic sphygmomanometers in acutely ill patients may not be the same as those recorded by ECG

Methods: heart rates recorded by an oximeter and an electronic sphygmomanometer were compared with electrocardiogram (ECG) heart rates measured on acutely ill medical patients.

Results: 1010 ECGs were performed on 217 patients while they were in the hospital. The bias between the oximeter and the ECG measured heart rate was -1.37 beats per minute (limits of agreement -22.6 to 19.9 beats per minute), and the bias between the sphygmomanometer and the ECG measured heart rate was -0.14 beats per minute (limits of agreement -22.2 to 21.9 beats per minute). Both devices failed to identify more than half the ECG recordings that awarded 3 NEWS points for heart rate.

Conclusion: Heart rates of acutely ill patients are not reliably measured by pulse oximeter or electronic sphygmomanometers.

[download-attachments container="div" style="list" display_icon="1" display_count="1" display__size="0"]

References

  1. Nannan Panday RS, Subbe CP, van Galen LS, Kellett J, Brabrand M, Nickel CH, Nanayakkara PWB. Changes in vital signs post discharge as a potential target for intervention to avoid readmission. Acute Med. 2018;17(2):77-82.
  2. Perera YS, Ranasinghe P, Adikari AM, Welivita WD, Perera WM, Wijesundara WM, Karunanayake SA, Constantine GR. The value of the Modified Early Warning Score and biochemical parameters as predictors of patient outcome in acute medical admissions a prospective study. Acute Med. 2011;10(3):126-32.
  3. Soong JTY, Soni N. Circulatory shock. Medicine 2013;41:64–9.
  4. Mok W, Wang W, Cooper S, Ang ENK, Liaw SY. Attitudes towards vital signs monitoring in the detection of clinical deterioration: scale development and survey of ward nurses. Int J Qual Health Care 2015;27:207–13.
  5. Fieselmann JF, Hendryx MS, Helms CM, Wakefield DS. Respiratory rate predicts cardiopulmonary arrest for internal medicine inpatients. J Gen Intern Med 1993;8: 354–60.
  6. Kellett J, Li M, Rasool S, Green GC, Seely A. Comparison of the heart and breathing rate of acutely ill medical patients recorded by nursing staff with those measured over 5 min by a piezoelectric belt and ECG monitor at the time of admission to hospital. Resuscitation. 2011;82:1381-1386.
  7. Opio MO, Kellett J for the Kitovu Hospital Study Group. How Well Are Pulses Measured? Practice-based evidence from an observational study of acutely ill medical patients during hospital admission. Am J Med. 2017 Jul;130(7):863.e13-863.
  8. Hollerbach AD, Sneed NV. Accuracy of radial pulse assessment by length of counting interval. Heart Lung. 1990;19:258-264.
  9. DeGowin EL, DeGowin RL. The heart and blood vessels. In: DeGowin EL, DeGowin RL, eds. Bedside Diagnostic Examination. 2nd ed. New York: Macmillan; 1969:337.
  10. Chung DK, Chung EK. A pulse deficit. Important finding in patients with artificial pacemakers. JAMA. 1972;219:1468-1469.
  11. Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1(8476): 307-310.
  12. Royal College of Physicians (RCP). National Early Warning Score (NEWS): Standardising the Assessment of Acute Illness Severity in the NHS. Report of a Working Party. London: RCP; 2012.
  13. Hamilton C, Stamey J. Using Bland-Altman to assess agreement between two medical devices – don’t for the confidence intervals! Journal of Clinical Monitoring and Computing 2007; 21:331–333
  14. World Medical Association. World Medical Association Declaration of Helsinki: ethical principles for medical research involving human subjects. JAMA 2013; 310:2191–4.
  15. Vandenbroucke JP, von Elm E, Altman DG, Gotzsche PC, Mulrow CD, Pocock SJ. et al. Strengthening the Reporting of Observational Studies in Epidemiology (STROBE): explanation and elaboration. Epidemiology. 2007;18(6):805-835.
  16. Myles P S. Using the Bland–Altman method to measure agreement with repeated measures. British Journal of Anaesthesia 2007;99:309–11.
  17. Iyriboz Y, Powers S, Morrow J, Ayers D, Landry G. Accuracy of pulse oximeters in estimating heart rate at rest and during exercise. Br J Sports Med. 1991 Sep; 25(3): 162–164.
  18. Wonka F. Oscillometric blood pressure measurement: description of the method used. Z. Kardiol. 1996;85:1-7.
  19. Beaubien E.R., Card C.M., Card S.E., Biem H.J., Wilson T.W. Accuracy of the Dinamap 1846 XT automated blood pressure monitor. J. Human Hypertension 2002;16:647-652.

Request Permissions

To request copyright permission to republish or share portions of our works, please visit Copyright Clearance Center’s (CCC) Marketplace website by clicking the button below:

Get Permission

How well are heart rates measured by pulse oximeters and electronic sphygmomanometers? Practice-based evidence from an observational study of acutely ill medical patients during hospital admission

19th September 2019
PMID: 31536051
Authors Affiliations
Mary Rimbi Medical Officer, Department of Medicine, Kitovu Hospital, Masaka, Uganda
Immaculate Nakitende Department of Medicine, Enrolled Nurse, Department of Medicine, Kitovu Hospital, Masaka, Uganda
Teopista Namujwiga Department of Medicine, Enrolled Midwife, Department of Medicine, Kitovu Hospital, Masaka, Uganda
John Kellett Department of Emergency Medicine, Hospital of South West Jutland, Esbjerg, Denmark

How well are heart rates measured by pulse oximeters and electronic sphygmomanometers? Practice-based evidence from an observational study of acutely ill medical patients during hospital admission

Cite this article as:

Rimbi M, Nakitende I, Namujwiga T, Kellett J. How well are heart rates measured by pulse oximeters and electronic sphygmomanometers? Practice-based evidence from an observational study of acutely ill medical patients during hospital admission. Acute Med. 2019;18(3):144-147. PMID: 31536051.

How well are heart rates measured by pulse oximeters and electronic sphygmomanometers? Practice-based evidence from an observational study of acutely ill medical patients during hospital admission

Share this article

Click the icon of the social media platform on which you would like to share this article.


Shopping Basket
Scroll to Top